Get peace of mind with SteriSense®
SteriSense is a reliable and cost efficient Electronic Bowie Dick Test that provides you with verification of critical sterilization parameters beyond that of chemical indicators.
SteriSense® Components and Functionalities
Process Challenge Device (PCD)
The PCD has been specially designed to reflect the reference method originally developed by Dr. J. Bowie and J. Dick in the 1960’s. The function of the PCD is to “challenge” the steam penetration of a steam sterilizer in accordance with EN ISO 11140-4. Between tests, the PCD needs to be cooled down to ambient temperature (approx. 90 minutes), which is why it is a big advantage that the device can easily be disassembled from the body (logger) and be changed with a spare. Due to the PCD design, SteriSense is also able to check for the presence of non-condensable gases (NCG).
SteriSense® Triple Sensor
1. Temperature sensor that measures inside the PCD
2. Temperature sensor that measures the ambient temperature in the sterilization chamber
3. Pressure sensor that measures the ambient pressure value within the chamber
SteriSense® Data Logger
The SteriSense data logger contains a battery with enough capacity to run 1,000 test cycles of 30 min. The data logger itself is based on Ellab’s 3rd generation of the well-known TrackSense® Pro data loggers.